Multi-level thermocouple

To the configurator

Multi-level thermocouple

type “T-M 15 (20,25,32,45,54,63) FHaKa“

Protection sleeve n~ measuring points Lmax./metres Mat.
1.5*0.15mm ≤3 3.0m VA/Inco*
2.0*0.2mm ≤7 3.0m VA/Inco*
2.5*0.2mm ≤11 3.0m VA/Inco*
3.0*0.2mm ≤19 10/3m VA/Inco*
3.2*0.2mm ≤23 10/3m VA/Inco*
3.5*0.2mm ≤28 10/3m VA/Inco*
4.0*0.5mm ≤26 25/25m VA/Inco*
4.5*0.5mm ≤36 25/25m VA/Inco*
5.0*0.5mm ≤48 25/25m VA/Inco*
6.35*0.89mm ≤64 25/25m VA/Inco*
7.0*1.0mm ≤77 25/25m VA/Inco*

specify according to table Ø diameter

Process connection

Document measuring point distribution here, or attach enquiry in separate file/sheet.

(Example: number and size or input screwed connections, number and design of required fittings such as plug connections or assembly aid etc.) optionally instead of connection box with transition dome forConnected Lemosa coupling directly in dome (box):

Example with flange
with union cap
with threaded rods
with plug connection

optionally instead of a connection box with transition dome inc. extension leads with/without assembled plug system:

isolation

Stufen-Thermoelement, Typ “T-M 15 (20,25,32,45,54,63) FHaKa“

In starrer Ausführung mit individuell konfektionierbarer Anschlussmöglichkeit

Set-up:

Thermocouples 1(or number n=x) Type “J“, “K“ und “N“, performance according to DIN EN 60584, Tolerance class 1 or 2, insulated or ground-welded, designed as MI cable, insulation made of highly compressed magnesium powder.

Installed in a protective fitting made of stainless steel or Inconel,e.g. with connection flange, neck tube and connection box made of stainless steel, aluminium or plastic with cable screw connections and installed series terminals, measurement transmitters are also possible.If required the cable screwed connections can also be replaced by various plug versions.

Other process connections or designs not listed here, can also be supplied. (For other options, please refer to the worksheet).

Specification of thermocouples and max. recommended temperature load for element type:

Fe-CuNi, “J“ NiCr-Ni, “K“ NiCrSi-NiSi, “N“ protection
600°C 600°C 600°C Stainless steel
600°C 800°C 800°C Alloy 600

For an improved response time the components can be connected under certain conditions with the inner walls of the protection sleeve; see special designs section or ask our technical customer service department.

Areas of application:

Chemicals: in vessels & reactors
Petrochemicals: in catalyst filling and distillation columns
Research: In laboratories and technical facilitiesfor media such as gases, vapours and liquids

Example type of wiring