Flanged thermocouple

  1. Home
  2. Thermocouples
  3. Flanged thermocouples
  4. Flanged thermocouple
To the configurator

FLANGED THERMOCOUPLE

FITTING WITH REPLACEABLE THERMOCOUPLE INSET

CALCULATION OF THE INSET LENGTH = PROTECTIVE TUBE LENGTH + EXTENSION PIPE LENGTH + 10MM

TEMPERATURE LOAD
OPERATING CONDITIONS

Structure of the fitting:

Structure Form 2F in accordance with DIN 43772; Protective tube according to DIN 43772, Diameter 11*2 mm or 14*2.5 mm of material stainless steel 1.4571; Flange DN 25 / PN 40 Form C of material 1.4571, Extension pipe of 11 mm (14 mm) diameter of material 1.4571, extension pipe length 82 mm, Connection head of light metal Form B in accordance with DIN EN 50446 with cable entry M20x1.5. (For variants please refer to the worksheet).

Structure of the gauge slide:

Thermocouples 1 (2) x Type “J”, “K” and “N”, Basic values in accordance with DIN EN 60584, tolerance class 1 or 2, Type “L” in accordance with DIN 43710, tolerance class 2, insulated or welded to the ground, inset tube of stainless steel with a 6.0 mm (8.0 mm) diameter; alternatively inset in flexible version as mineral insulated sheathed thermocouple inset of 6 or 8 mm diameter of stainless steel or Inconel material. Shelf plate with connection socket.
(For variants please refer to the worksheet).

Types of wiring:

Menu