Immersion resistance thermometer

Fittings with replaceable resistance thermometer inset

Structure of the fitting:

Structure Form 1 according to DIN 43772 with protective tube diameter of 9 – 15 mm of stainless steel material 1.4571 or teflonised, on request with a stop flange according to DIN 43734 or movable clamp connection; refer to catalogue group “Components”, Connection head of light metal alloy Form A or Form B in accordance with DIN EN 50446 with cable entry M20x1.5.
(For variants please refer to the worksheet).

Structure of the inset:

According to DIN43735; Precision resistor single or dual, Type Pt-100 as per IEC 60751 in Tolerance class B or A. Connected in 2-, 3- or 4-wire circuit from precision resistor, installed in a rigid inset pipe of stainless steel material or as sheath resistance thermometer sensor device in flexible version with a diameter or 6 or 8 mm. Shelf plate with connection socket.